3.3 4-20 mA Wiring
Connections of channel 1 are on terminals 6 and 7. For channel 2 on
terminals 2 and 3. Observe indicated polarity.
3.4 0-50 mV Wiring
Connections of channel 1 are on terminals 6 and 7. For channel 2 on
terminals 2 and 3. Observe indicated polarity.
3.5 0-10 V Wiring
Linear voltage signals from 0 to 10 V can only be connected to
channel 1, terminals 5 and 6. Observe indicated polarity.
3.6 Dry Contact Digital Input Wiring
Digital signals are connected to channel 3, terminals 1 and 2.
The contact connected to digital input must not have any electrical voltage.